This refurbished Agilent 1100 HPLC G1311A Quaternary Pump Quat Pump is fully tested in working condition
Including:
-
- Agilent 1100 HPLC G1311A Quaternary Pump Quat Pump
- Installation guide and user documentation
- Power and communication cables
- 90 Days Warranty
Specification | Details |
---|---|
Hydraulic system | Dual plunger in series pump with proprietary servo-controlled variable stroke drive, floating plungers, and active inlet valve |
Setable flow range | 0.001 – 10 ml/min, in 0.001 ml/min increments |
Flow range | 0.2 – 10.0 ml/min |
Flow precision | < 0.3 % RSD (typically 0.15 %), based on retention time, at 1 ml/min |
Pressure (0 – 5 ml/min) | 0 – 40 MPa (0 – 400 bar, 0 – 5880 psi) |
Pressure (5 – 10 ml/min) | 0 – 20 MPa (0 – 200 bar, 0 – 2950 psi) |
Pressure pulsation | < 2 % amplitude (typically < 1 %), at 1 ml/min isopropanol, at all pressures > 1 MPa (10 bar) |
Compressibility compensation | User-selectable, based on mobile phase compressibility |
Recommended pH range | 1.0 – 12.5; solvents with pH > 2.3 should not contain acids which attack stainless steel |
Gradient formation | Low pressure quaternary mixing/gradient capability using proprietary high-speed proportioning valve |
Delay volume | 800 – 1100 µl, dependent on back pressure |
Composition range | 0 – 95 % or 5 – 100 %, user selectable |
Composition precision | < 0.2 % SD, at 0.2 and 1 ml/min |