Description

This refurbished Agilent 1260 Infinity G1314F VWD Variable Wavelength Detector is fully tested in working condition
Including:

    • Agilent 1260 Infinity G1314F VWD Variable Wavelength Detector
    • Installation guide and user documentation
    • Power and communication cables
    • 90 Days Warranty
Type Specification
Detection type Double-beam photometer
Light source Deuterium lamp
Wavelength range 190 – 600 nm
Short-term noise ± 0.25 × 10-5 AU at 230 nm
Drift < 1 × 10-4 AU/h at 230 nm
Linearity > 2.5 AU (5 %) at 265 nm
Wavelength accuracy ± 1 nm
Maximum data rate 80 Hz
Bandwidth 6.5 nm typical
Flow cells Standard: 14 μL volume, 10 mm cell path length, 40 bar (588 psi) max pressure
High pressure: 14 μL volume, 10 mm cell path length, 400 bar (5880 psi) max pressure
Micro: 2 μL volume, 3 mm cell path length, 120 bar (1760 psi) max pressure
Semi-micro: 5 μL volume, 6 mm cell path length, 40 bar (588 psi) max pressure
Electronic Temperature Control (ETC) Improves baseline stability in unstable environments
Control and data evaluation Agilent ChemStation B.04.02 SP2 or above
Instant Pilot (G4208A) with firmware B.02.11 or above
Time programmable Wavelength, Reference and Sample scan, balance, test, lamp on/off
Spectral tools Stop-flow wavelength scan
Analog outputs Recorder / integrator: 100 mV or 1 V
Output range: 0.001 – 2 AU, one output
Communications LAN and integrated on main board
Controller-area network (CAN)
RS-232C
APG Remote: ready, start, stop and shutdown signals
Safety and maintenance Extensive troubleshooting and maintenance support via Instant Pilot, Agilent Lab Advisor, and Chromatography Data System
Safety features include leak detection, safe leak handling, leak output signal for shutdown, and low-voltage areas in major maintenance zones
GLP features Early Maintenance Feedback (EMF) for tracking lamp burn time with user-set limits and messages
Electronic maintenance and error records
Verification of wavelength accuracy using built-in holmium oxide filter
RFID for electronic records of flow cell and UV lamp conditions (path length, volume, product number, serial number, test passed, usage)
Housing All materials recyclable