This refurbished Waters ACQUITY Arc QSM-R P/N: 186017000 Quaternary Solvent Management-R is fully tested and in working condition
Including:
- Waters ACQUITY Arc QSM-R P/N: 186017000 Quaternary Solvent Management-R
- Installation guide and user documentation
- Power and communication cables
- 90 Days Warranty
Overview:
The Waters™ ACQUITY™ Arc™ System is the only available LC system with regulatory compliant-ready hardware and software
for both HPLC and UHPLC separations in one instrument. With an expanded range of column manager options enabling up to
15-column switching from 4 °C to 90 °C and Arc Multi-flow path™ technology, the ACQUITY Arc System allows you to generate
new methods and modernize legacy HPLC methods faster than ever before.
Specifications:
QUATERNARY SOLVENT MANAGER-R | |
---|---|
Solvent Capacity | Blend up to four solvents in any combination (standard); total capacity of nine solvents with integrated solvent select valve (optional) |
Number of Fluidic Paths | Three (Path 1, Path 2, and waste), with Arc Multi-flow path technology (standard) |
Dwell Volume Selection | Automated with Arc Multi-flow path technology |
Solvent Conditioning | Integrated vacuum degassing, four chambers |
Gradient Formation | Low-pressure mixing, quaternary gradient |
Gradient Profiles | 11 gradient curves [including linear, step (2), concave (4), and convex (4)] |
Check Valves | Passive check valves |
Flow Accuracy | ±1.0% at 0.5, 3.0, and 5.0 mL/min |
Flow Precision | ≤0.075% RSD or ±0.020 min SD, whichever is greater, based on six replicates [60:40 water:methanol pre-mixed; 1.5 mL/min; alkylphenone mix; 24.0 µL injection volume; CORTECS™ C18 2.7 µm, 4.6 × 50 mm; 35 °C; UV @254 nm] |
Composition Ripple | ≤0.5 mAU [mobile phase containing 0.1% TFA in water/acetonitrile; 1.5 mL/min; CORTECS C18 2.7 µm, 4.6 × 50 mm; 35 °C; UV @214 nm] |
Composition Accuracy | ±0.5% absolute (full scale) from 5 to 95%; 0.5 to 5.0 mL/min [methanol; methanol with 5.0 mg/mL caffeine step gradient; UV @273 nm] |
Composition Precision | ±0.15% RSD or 0.04 min SD, whichever is greater based on six replicate injections [60:40 water:methanol via Auto•Blend™ Technology; 0.5 mL/min; alkylphenone mix; 24.0 µL injection volume; CORTECS C18 2.7 µm, 4.6 × 50 mm; 35 °C; UV @254 nm] |
Compressibility Compensation | Automatic and continuous |
Priming | Wet priming can run at flow rates up to 10 mL/min |
Pump Seal Wash | Standard |
Primary Wetted Materials | 316L stainless steel, PPS, fluoropolymer, UHMWPE blend, sapphire, ruby, zirconia, DLC, PEEK and PEEK blend, titanium alloy |