This refurbished Waters Acquity UPLC Sample Manager SM P/N:186015006 is fully tested in working condition
Including:
-
-
- Waters Acquity UPLC Sample Manager SM P/N:186015006
- Installation guide and user documentation
- Power and communication cables
- 90 Days Warranty
-
-
Overviews:
The ACQUITY UPLC sample manager injects the samples it draws from microtiter plates or vials onto the chromatographic column. A locating mechanism uses a probe to access sample locations and draw sample from them. In the needle overfill load-ahead mode, the sample manager can perform an injection in approximately 15 seconds. The first injection requires additional overhead time. The sample manager accepts standard ANSI/SBS footprint plates, 5.03 ±0.02 inch × 3.365 ±0.02 inch, that conform to ANSI standards (maximum height = 2.2 inches, including covers). You can program any combination of these plates and vial holders for automated sample processing. Samples are loaded into the sample manager via the front door or the optional sample organizer, which transfers samples back and forth between the two instruments. The sample manager can maintain samples at any temperature between 4 and 40 ºC (39.2 to 104 ºF) in 25 ºC (77 ºF) or less ambient conditions.